ME 26 ACE-INC
for transducer with integrated electronic : accelerometer, inclinometer ...

Zero adjustment

 
Voltage measuring : Input 2.5 V +/- 2V
Transducer power supply : 5 Vdc or 26 Vdc
Accuracy and stability
Analogic output : ± 10V or 0 -20 mA or 4 - 20 mA
Zero and gain adjustment
Electronic calibration simulator
Bandwidth from 1 kHz to 20 kHz
Easy adjustment 
Span adjustment
Calibration
Scale adjustment

Compatible with : 
- 2680C  multi-channels instrument, 
-
ME52 rail DIN mono-channel
- ME261B IP65 multi-channels.

 

APPLICATION AND CONNECTION

 

TECHNICAL FEATURES

Power supply (Without output charged) 24 Vdc ± 2V, 25 mA
Transducer power supply 5 Vdc or 26 Vdc (other on request)
Input (admissible maximum ± 25V) Input voltage range : 2.5 V +/- 2 V
Another range on request
Output voltage Maximum output voltage : ± 10V
Maximum output current : 5 mA
Maximum output impedance : 50 Ohm
Current output (option) 0 to 20 mA or 4 to 20 mA
Zero adjustment 2.5 V +/- 0.5 V
Calibration simulator 1V
Bandwidth (at -3dB) 1000 Hz (other on request)
Linearity 10E-04
Maximum zero drift in the input 2 µV/°C
Common mode rejection 100 dB
Power supply variations rejection (± 2V) 100 dB
Maximum noise see to input (White noise) 0.2 mV peak to peak
Maximum noise see to input (with ME 26 ISO) 10mV peak to peak
Operating temperature 0°C to 50°C
Storage temperature -40°C to +70°C
Protection Against short circuits
Option  RAL26EU 100 x 160 mm Europe  adapter DIN 41612 connector
Galvanic Insulation Option with ME26ISO with extension card ME 26 ISO
Input / Output Insulation : 500 V
Input / Power supply Insulation:   500 V