ME26AJ   OR  ME26AS

ANALOGIC OUTPUT ± 10V FOR STRAIN GAGES

Zero adjustment

Load, pressure, torque...
For any transducer with strain gages, (classic or piezoresistant)
Accuracy and stability
Transducer power supply
Analogic output ± 10V or 0/20 mA or 4/20 mA
Zero and gain adjustment
Electronic calibration simulator
Easy adjustment
Span adjustment
Calibration
Scale adjustment
Compatible with : 
- 2680C  multi-channels instrument, 
-
ME52 rail DIN mono-channel
- ME261B IP65 multi-channels.

APPLICATION AND CONNECTION

 

TECHNICAL FEATURES

Bridge power supply voltage, given by the board 10 V (other value on request)
Bridge impedance from 350 Ohm to 1 KOhm.
(features given for 350 Ohm)
Sensitivity of the transducer from 0.5mV/V to 3.5 mV/V for the reference ME26AJ
and from 5 to 35 mV/V for the reference ME26AS
.
Electronic calibration simulator 1mV/V for ME26AJ
and 10mV/V for ME26AS
Adjustable zero and gain by trimmer Adjustable zero +/- 1mV/V for ME26AJ
and +/- 10mV/V for ME26AS
Linearity 1. 10-4
Analogic output ± 10V / 0V analogic or 0-20 mA or 4-20mA / Ground
Output voltage bandwidth 1KHz (other value on request, from 100 Hz up to 10 KHz).
Protection against short circuits.
Power supply 24 Vdc ± 2V
Consumption with 24 Vdc 40mA
Option RAL26EU 100 x 160 mm Europe   adapter DIN 41612 connector