ME26DP
ANALOGIC OUTPUT FOR POTENTIOMETRIC TRANSDUCER
![]() |
Zero adjustment |
|
||||||||||||||||
| Span adjustment | ||||||||||||||||||
| Calibration | ||||||||||||||||||
| Scale adjustment | ||||||||||||||||||
|
Compatible with : - 2680C multi-channels instrument, - ME52 rail DIN mono-channel - ME261B IP65 multi-channels. |
||||||||||||||||||
APPLICATION AND CONNECTION

TECHNICAL FEATURES
| Power supply for the sensor | 5 ± 0.1 V Max Drift: 30 ppm/°C |
| Impedance | 200 Ohm to 1 MOhm |
| Output voltage | Maximum output voltage : ± 10 V Maximum output current : 5 mA Maximum output impedance : 50 Ohm |
| Output current (option) | 0 to 20 mA or 4 to 20 mA |
| Input sensitivity | Range III : from 0.2 to 0.75 V/V Range II : from 0.75 to 3.4 V/V Range I : Unusable for DP |
| Fixed zero adjustment at ± 5% | Range III : ± 1.1 V/V Range II : ± 0.25 V/V Range I : Unusable for DP |
| Adjustable zero | Range III : ± 1.1 V/V Range II : ± 0.25 V/V Range I : Unusable for DP |
| Calibration simulator | 0.5V/V |
| Bandwidth (at -3dB) (another value on request) |
Range III : 1 kHz Range II : 1.7 kHz Range I : Unusable for DP Option : wide bandwidth 10 KHz |
| Rising time | < 250 µs |
| Linearity | 10-4 |
| Zero maximum drift in the input | < 2 µV/°C |
| Common mode rejection | 100 dB |
| Power supply variations rejection (board in the 2680 instrument) |
120 dB |
| Maximum noise in the input (White noise) | < 3m V eff |
| Maximum noise in the input (with ME 26 ISO) | < 150m V eff |
| Operating temperature | 0°C to 50°C |
| Storage temperature | -40°C to +70°C |
| Power supply | 24V +/ 2V continuous, 50 mA |
| Protection | against the short circuits |
| Option: RAL26EU | 100 x 160 mm Europe adapter DIN 41612 connector |