ME26DP

ANALOGIC OUTPUT  FOR POTENTIOMETRIC TRANSDUCER

Zero adjustment

Displacement, Position ...
For all potentiometric transducers
Accuracy and stability
Transducer power supply
Analogic output ± 10V or 0 - 20 mA or 4 - 20 mA
Zero and gain adjustment
Electronic calibration simulator 0.5 V/V
Easy adjustment 
Span adjustment
Calibration
Scale adjustment
Compatible with : 
- 2680C  multi-channels instrument, 
-
ME52 rail DIN mono-channel
- ME261B IP65 multi-channels.

APPLICATION AND CONNECTION

 

TECHNICAL FEATURES

Power supply for the sensor 5 ± 0.1 V
Max Drift: 30 ppm/°C
Impedance 200 Ohm  to 1 MOhm
Output voltage Maximum output voltage : ± 10 V
Maximum output current : 5 mA
Maximum output impedance : 50 Ohm
Output current (option) 0 to 20 mA or 4 to 20 mA
Input sensitivity Range III : from 0.2 to 0.75 V/V
Range II : from 0.75 to 3.4 V/V
Range I : Unusable for DP
Fixed zero adjustment at ± 5% Range III : ± 1.1 V/V
Range II : ± 0.25 V/V
Range I : Unusable for DP
Adjustable zero Range III : ± 1.1 V/V
Range II : ± 0.25 V/V
Range I : Unusable for DP
Calibration simulator 0.5V/V
Bandwidth (at -3dB)
(another value on request)
Range III :  1 kHz
Range II :  1.7 kHz
Range I : Unusable for DP
Option : wide bandwidth 10 KHz
Rising time < 250 µs
Linearity 10-4
Zero maximum drift in the input < 2 µV/°C
Common mode rejection 100 dB
Power supply variations rejection
(board in the 2680 instrument)
120 dB
Maximum noise in the input (White noise) < 3m V eff
Maximum noise in the input (with ME 26 ISO) < 150m V eff
Operating temperature 0°C to 50°C
Storage temperature -40°C to +70°C
Power supply 24V +/ 2V continuous, 50 mA
Protection against the short circuits
Option: RAL26EU 100 x 160 mm Europe  adapter DIN 41612 connector