ME 26 PIEZO = ME 26 ACP Module + ME 26 UV2 Card
ANALOGIC OUTPUT ± 10V FOR PIEZZOELECTRIC TRANSDUCERS (QUARTZ)
ME 26UV2 Card![]() ACP Module |
Adjustable zero |
|
||||||||||||||||||
| Reset by pusing switch or/and at distance | ||||||||||||||||||||
| Adjustable gain | ||||||||||||||||||||
| Calibration | ||||||||||||||||||||
| Scale adjustment | ||||||||||||||||||||
| Conditioner board compatible with the box referenced : 2680 C | ||||||||||||||||||||
APPLICATION AND CONNECTION

TECHNICAL FEATURES
ME 26 ACP MODULE |
|
| Output sensitivity | 0.45 mV / pC |
| Output | +/- 9 V for +/- 20 000 pC |
| Linearity | < 0.1 % |
| Accuracy | < 1 % |
| Output maximum current | 0.1 mA |
| Compatible output with the amplification board | ME 26 UV2 |
| Reset Reset typical number Time mini for Reset operation |
by internal
relay 200. 106 cycles (MTBF) 500 ms |
| Bandwidth | from 0.002 Hz to 10 KHz (-3dB) |
| Power supply Consumption |
from 24
VDC to 30 VDC < 30 mA |
| Transducer Input connector | BNC Embase |
| Voltage output connector | 7 male dowels Amphenol plug |
| Case dimensions | 50 x 50 x 15 mm |
| Case type | EMI/RFI armored |
ME26UV2 CARD |
|
| Voltage Amplifier | for ME26ACP module |
| Sensitivity | Adjustable from 200
pC to 20.000 pC in 3 ranges |
| ME26ACP module power supply | 26 VDC |
| Calibration simulator | 1000 pC |
| Analogic output | +/- 10 V |
| Bandwidth | from 0 to 5 KHz |
| Charge amplifier reset | by pushing switch |
| Option MV2 | Peak memory |
| Option RAL26EU | 100 x 160 mm Europe adapter DIN 41612 connector |