ME 26 TC
ANALOGIC OUTPUT FOR J, K, T ... THERMOCOUPLE
![]() |
|
|||||||||||||||
| Calibration | ||||||||||||||||
|
Compatible with : - 2680C multi-channels instrument, - ME52 rail DIN mono-channel - ME261B IP65 multi-channels. |
||||||||||||||||
APPLICATION AND CONNECTION
TECHNICAL FEATURES
| Power supply (Without charge in the output) |
26 V continuous, 36 mA for an +/- 10 V
output 24 V for 4 - 20 mA output |
| Inputs | J, K, T ... Thermocouples (to specify) |
| Weld compensation | from - 10°C to + 50°C |
| Output voltage | Maximum output voltage : ± 10 V Maximum output current : 5 mA Maximum output impedance : 50W |
| Output current (option) | 0 - 20 mA or 4 - 20 mA |
| Measuring ranges | Range I : from +/- 200°C Range II : from 0 - 1.000 °C Range III : to define |
| Zero adjustment | Adjusted in our laboratory |
| Calibration simulator | 100 °C or 1.000 °C (other on request) |
| Bandwidth (at -3dB) | 10 Hz (until 1.000 Hz on request) |
| Maximum zero drift in the input | 2 µV/°C |
| Common mode rejection | 100 dB |
| Power supply variations rejection (± 2V) | 100 dB |
| Maximum noise in the input (White noise) | 0.2 mV peak to peak |
| Maximum noise in the input (with ME 26 ISO) |
10 mV peak to peak |
| Operating temperature | 0°C to 50°C |
| Storage temperature | -40°C to +70°C |
| Protection | Against short circuits |
| Option RAL26EU | 100 x 160 mm Europe adapter DIN 41612 connector |
| Option: ME26ISO (extension board). | 500 V input / output insulation 500 V input / Power supply insulation |